Gima H100 Instrucciones Para El Uso página 79

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

PRO BLEM S
The autoclave does not
sw itch on
START
and
H EATING leds do not
light on w hen turning the
program m er knob
H eating led lights on but
tem perature does not
increase
W hile the autoclave
w orking,
the ALARM
led lights on (6)
If bad working goes on, call technical assistance and communicate the model of the unit and its
serial number, which is on the label, on the back of the unit, or on the warranty.
Autoclave GIMA H100 automatic
CAUSES
cut-out switch of the unit and /or
differential cut-out are O FF.
lack of tension in the outlet where
the unit is connected
M ains fuses are interrupted
the plug of current is not connected
check if the alarm led is on
electric resistance is burnt
electronic card is faulty
the safety thermostat has worked
is
position the switches ON
check the cause of lack of tension in the
outlet and find a remedy
Replace them with new ones of the same
value
(see
chapter 12 "M aintenance and cleaning"
on page no. 79)
connect the plug.
If ALARM led is on you must push the
safety thermostat button which is on the
back of the autoclave
call
technical
autoclave back.
call
technical
autoclave back
Turn the autoclave off
W ait till the autoclave is
min.)
Unscrew the black
back side of
no. 10 on page no. 65).
W ith
a
button that is placed under the cap.
Heck if the
completely covered with water.
Repeat sterilizing cycle.
- 75 -
REM EDIES
"w ays
of
replacem ent",
assistance
or
send
servicing
or
send
cold (10 / 15
cap situated on the
the autoclave (see detail
sharpened object
push the red
temperature control probe is
the
the

Hide quick links:

Publicidad

Capítulos

Tabla de contenido
loading

Productos relacionados para Gima H100

Tabla de contenido