9 Program Values

9
Program Values
9.1

Program Values world-wide (except USA and Asia)

Index 30
Program
Program
duration
approx.
P1
21s
P6.1 + P6.2
21s + 21s
P11
18s
Possible kV/mA combinations – can be selected manually
60
62
64
66
10
10
10
10
Film and Processing Notes
30
Exposure
Factory-programmed values
time
with a film density of 03
11.3s
66/10
70/10
7.8s
68/10
72/10
11,3s
66/10
70/10
68
70
72
74
10
10
10
10
Operating Instructions ORTHOPHOS 3
Intensifying Screen Kodak Lanex Regular (sensitive to
green) with film Kodak T-Mat G, Agfa Ortholux. Differ-
ences in film density depending on film and development
tolerances can be compensated by changing the posi-
tion of the density switch (see page 28).
From 03 to 02 when the density is too high, from 03 to 04
when the density is too low.
Freely programmed values or
values with other film density: ....
74/10
78/10
76/10
80/10
74/10
78/10
These values serve only as user reference times.
76
78
80
10
10
10
The programmed values apply to Kodak T-Mat G and
Agfa Ortholux films in combination with the Kodak Lanex
Regular intensifying screen.
Process according to the manufacturer of the developing
equipment's guidelines or follow instructions on the
chemicals.