6 - Exploring the Generator Capabilities
Setting-up an arbitrary wave signal
6. EXPLORING THE GENERATOR CAPABILITIES
In the following examples only the parameter settings are described, together with the related key
names. The resultant output waveforms are shown, along with the sync or trigger waveform where
relevant. Output amplitude and offset settings are examples only and need not be followed.
Setting-up an arbitrary wave signal
Start with the instrument returned to Default Settings.
MENU
Waveform
MENU
Arbitrary waveform
Parameter
Wave selection
Frequency
Amplitude
Offset
Parameter
Output State
32
HARD KEY NAME
Waves
Soft-key Name
Arb
Soft-key Name
Waves
-
Ampl
Offset
HARD KEY NAME
Output1
Setting
Sinc
10kHz
2V
5mVdc
Setting
On
TGF4000 Series Quick Start Guide