BS series weighing scales user's manual

7.8 Change capacity

Before this operate, please short K4
After enter setting mode, display shows "F0 CAL", press ZERO key six times, display
shows "F6 CAP", press TARE key, display show "pin---", waiting for input password,
press TARE key and ZERO key to input password (TARE key for increase value,
ZERO key for move active digit), default password is "315", after input correct
password, display will show capacity, press TARE key to change(option: 3kg, 6kg,
15kg, 25kg), press ZERO key to sure
7.9 Auto zero range
After enter setting mode, display shows "F0 CAL", press ZERO key seven times,
display shows "F7 Azn", press TARE key, the parameters will set the auto zero
range.(option: 0.5d/1d/2d/4d)
7.10 Speed
After enter setting mode, display shows "F0 CAL", press ZERO key seven times,
display shows "F8 SPD", press TARE key, the
parameters will
set
weighing
speed.( option: 7.5/15/30 times).

7.11 Reset

After enter setting mode, display shows "F0 CAL", press ZERO key seven times,
display shows "F9 RST", press TARE key, the parameters will set back to the default
value(factory mode).
7.12 Gravity
To set gravity, press TARE/UNIT and ZERO key together, when the scale in the self
test. it will show GRAV, input the gravity value, then press ZERO key to make sure.