VWR High Performance Guia De Inicio Rapido página 11

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Single models
Volume
Model
range (µL)
2µl
0.2 - 2
10µl
1 - 10
20µl
2 - 20
100µl
10 - 100
200µl
20 - 200
1000µl
100 - 1 000
5000µl
500 - 5 000
10mL
1000 - 10000
The data given in the tables are achieved with VWR
standard length series pipette tips. If you are using
VWR 10 µL extended length series pipette tips, you
will need to recalibrate your pipettor to comply with
specifications.
SPECIFICATIONS - Error limits
Volume
Systematic error
(µL)
(µL)
0.2
± 0.026 ± 13.2
1
± 0.030 ± 3.0
2
± 0.033 ± 1.7
1
± 0.035 ± 3.5
5
± 0.083 ± 1.6
10
± 0.110 ± 1.1
2
± 0.11
10
± 0.12
20
± 0.20
10
± 0.39
50
± 0.44
100
± 0.80
20
± 0.55
100
± 0.88
200
± 1.60
100
± 3.3
500
± 4.4
1000
± 8.0
500
± 13
2500
± 17
5000
± 33
1000
± 33
5000
± 44
10000
± 60
QUICK START GUIDE
Random error
(%)
(µL)
≤ 0.013 ≤ 6.6
≤ 0.014 ≤ 1.4
≤ 0.015 ≤ 0.8
≤ 0.013 ≤ 1.3
≤ 0.033 ≤ 0.6
≤ 0.044 ≤ 0.4
≤ 0.033 ≤ 1.7
± 5.5
≤ 0.055 ≤ 0.5
± 1.2
≤ 0.066 ≤ 0.3
± 1.0
≤ 0.11
± 3.9
≤ 0.13
± 0.8
≤ 0.17
± 0.8
≤ 0.22
± 2.8
≤ 0.28
± 0.8
≤ 0.33
± 0.8
≤ 0.7
± 3.3
≤ 1.1
± 0.8
≤ 1.7
± 0.8
≤ 3.3
± 2.6
≤ 5.5
± 0.6
≤ 8.8
± 0.7
≤ 6.6
± 3.3
≤ 11.0
± 0.8
≤ 17.6
± 0.6
(%)
≤ 1.1
≤ 0.2
≤ 0.2
≤ 1.1
≤ 0.2
≤ 0.2
≤ 0.7
≤ 0.2
≤ 0.2
≤ 0.7
≤ 0.2
≤ 0.2
≤ 0.7
≤ 0.2
≤ 0.2
11
EN

Publicidad

Tabla de contenido
loading

Tabla de contenido