Aspirate
Press the push-button to the first stop (this
corresponds to the set volume of liquid).
Hold the pipettor vertically and immerse the tip in
the liquid (see immersion depth table, in the Online
User Manual).
Release the push-button slowly and smoothly
(to top position) to aspirate the set volume of liquid.
Wait one second; then withdraw the pipette tip from
the liquid.
You may wipe any drop lets away from the outside
of the tip using a medical wipe, however if you do so
take care to avoid touching the tip's orifice.
/ For the multichannel models, use a reagent reservoir.
Dispense
Place the end of the tip against the inside wall of the
recipient vessel (at an angle of 10° to 40°).
Press the push-button slowly and smoothly to the
first stop.
Wait for at least a second, then press the push-button
to the second stop to expel any residual liquid from
the tip.
Keep the push-button pressed fully down and (while
removing the pipettor) draw the tip along the inside
surface of the vessel.
Release the push-button, smoothly. Eject the tip by
pressing firmly on the tip-ejector button.
/ For the multichannel models, use a reagent reservoir.