ELECTROMAGNETIC COMPATIBILITY
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF
communications equipment and this monitor
This monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of this monitor can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and this monitor as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output
power of
transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Mode of operation: Continuous operation
Degree of protection: Type B applied part
IP Classification :IPX0
Software version: V01
Device classification: Internally Powered ME Equipment
Table 4
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d =
1
2 .
0,12
0,38
1,2
3,8
12
m
80 MHz to 800 MHz
d =
1
2 .
0,12
0,38
1,2
3,8
12
800 MHz to 2,5 GHz
d = 2.3
0,23
0,73
2,3
7,3
23