Overview; Structure And Function; Recommended Fields Of Application - Xylem SI Analytics LF 413T IDS Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

LF 413T IDS
Structure
Automatic sensor
recognition
ba77043d02
10/2015
1

Overview

1.1

Structure and function

1, 2 3
1
Voltage electrode (inside, 2x)
2
Current electrode (ring, 2x)
3
Temperature sensor in graphite enclosure
4
Shaft
5
Connecting head with active electronics
The sensor electronics with the stored sensor data is in the connecting head.
The data include, among other things, the sensor type and series number.
With each calibration, the calibration data is written in the sensor and the
calibration history is recorded. The data is recalled by the meter when the
sensor is connected and is used for measurement and for measured value
documentation. Storing the calibration data in the sensor ensures that the
correct cell constant is automatically used if the sensor is operated with
several meters.
The digital transmission technique guarantees the failure-free
communication with the meter even with long connection cables. If the sensor
firmware is enhanced by SI Analytics, it can be updated via the meter.
1.2

Recommended fields of application

 On site measurements in rivers, lakes and wastewater
 Fish farming
 Ground water measurements
 Applications in water laboratories
Due to the use of high performance PEEK material for the connection
technology has an improved chemical resistance, especially in acid media
(eg. As electroplating baths).
NOTE
Concentrated or strong oxidizing acids and organic solvents can
damage or destroy the sensor.
4
Overview
5
5

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Si analytics lf 413t 3m ids

Tabla de contenido