Measuring / Operation

Scope of delivery
Preparing the
sensor for
measurement
Outside cleaning
Aging of the
conductivity
measuring cell
Disposal
6
2
Measuring / Operation
 Conductivity measuring cellLF 413T IDS
 Operating manual
Connect the sensor to the meter.
The sensor is immediately ready to measure.
3

Cleaning

NOTE
Prior to cleaning, disconnect the sensor from the meter.
We recommend to clean the sensor thoroughly, especially before measuring
low conductivity values.
Contamination
Lime sediments
Fat/oil
After cleaning, thoroughly rinse with deionized water and recalibrate if
necessary.
Normally, the conductivity measuring cell does not age. Special measuring
media (e.g. strong acids and bases, organic solvents) or temperatures that
are too high may considerably reduce its lifetime or lead to damage. The
warranty does not cover failure caused by measuring conditions and
mechanical damage.
We recommend to dispose of the measuring cell as electronic waste.
Cleaning procedure
Immerse in acetic acid for 5 minutes (volume
share = 10 %)
Clean with warm water that contains wash-
ing-up liquid
LF 413T IDS
ba77043d02
10/2015