CALIBRATION

This scale has been factory calibrated, and does not require calibration prior to use. If required, the scale can be
calibrated following the procedure below.
Note: Only weights certified and traceable to national standards are to be used for calibration procedures.
KL Model Calibration
KL model scales can be calibrated in pounds or kilograms. The default calibration weight unit is pounds. If you prefer
to calibrate in kilograms you must first follow this procedure to change the calibration weight unit.
Caution: Please make sure you carefully and precisely follow the directions below as you will be in the
PROGRAM MODE of your scale. Any modification outside of what is listed below may greatly affect the
functionality of your scale.
To Change the Calibration Weight Mode to KG
1. Press the ON/OFF button to power on the scale. As soon as you turn on the scale, press the ZERO button once,
the KG/LB button twice, and then the ZERO button once again. The scale will display "CAL".
2. Press the
once and "PROG" will show on the display.
3. Press the ENTER button and OP1-0 or OP1-1 will be displayed.
4. Press the UNITS button to toggle between OP1-0 and OP1-1. Select OP1-0 to set your calibration weight in
pounds. Select OP1-1 to set your calibration weight to kilograms.
5. After making your selection, press the
6. Press the ENTER button, the scale will read "START", followed by a version code and then return to zero.
7. Turn the scale off. Then begin the calibration procedure below.
KG Model Calibration
KG models can only be calibrated in kilograms. Follow the calibration procedure below.
Calibration Procedure for KL and KG Models
The calibration weight must be no less than 60 lb (28 kg) and no more than 300 lb /150 kg. For optimum
accuracy, Health o meter Professional Scales recommends calibrating at 300 lb / 150 kg.
OPERATOR ACTION
Turn on the scale and while "START' is displayed, press the following keys in this
sequence:
ZERO, KG/LB, KG/LB, ZERO
Note: You must press these keys while "START" is displayed, before the display
shows zero.
Press ENTER. "Load" and "300.0" or "150.0" will appear on the display
Press ENTER to calibrate with 300 lb or 150 kg. To calibrate with a different amount,
press BMI. Use the arrows to change the flashing digit, BMI to move to the next
digit, until you reach the required weight (no less than 60lb and no more than 300lb).
When the required calibration weight is reached, press ENTER. "Clear" will appear
on the display.
Clear the platform of any weight and press ENTER. Dashes will appear while the
scale calibrates zero.
Wait without touching or shaking the scale until the zero calibration process is
finished. The display will then prompt for the calibration weight to be loaded.
Load the platform with the required weight and press ENTER. Dashes will appear on
the display as the scale calibrates.
When calibration is complete "DONE/CLEAR/XXX" will appear on the display.
Remove all weight from the platform. When all weight is cleared the scale will reboot
and "StArt" will appear on the display followed by the software version and then
return to "0.0".
Refer to the next page should you experience any errors during the calibration procedure.
P/N UM597KL_KG 599KL_KG 752KL_KG REV20191121_Web
until display reads "DONE"
18
DISPLAY
"CAL"
"Load/300.0" or
"Load/150.0"
"CLEAR"
"-----------"
"Put XXX.X"
"-----------"
"DONE/CLEAR/XXX"
"StArt"; numbers; "0.0"