6.3
Assembly of the
optional foot switch
6.4a
Assembly of
the REUSABlE
collection system
positions for jars
(depending on the
configuration):
1.
1. If you use a DISPOSABlE collection system proceed
with step 6.4b
2.
2.2
3.
3.2
3.1
4.
5.
1.1 Connect the optional foot
switch to the pump by
plugging in the plug.
1.2 Test the correct function-
ning of the foot switch.
2.1 Attach mechanical over-
flow protection to lid.
2.2 Pull gently downwards to
make sure it is
2.1
open / deactivated.
3.1 Attach the lid to the jar and
lock with the two lid
clamps.
3.2 Attach the jar to the carry-
ing handle (portable version
3.1
only) or to the optional
clampholder on the rail (see
step 5.4).
4.1 Connect tubing from the
Safety Set to the lid of the
jar (vacuum port).
5.1 Connect patient tubing to
the lid of the jar (patient
port).
Preparation for use
| 15