6.4b
Assembly of
the DISPOSABlE
collection system
positions for jars
(depending on the
configuration):
6.5
Install equipotential
bonding
to bus bar
16 |
Preparation for use
1. If you use a REUSABlE collection system proceed with
step 6.4a
2.
3.
4.
A B
5.
1.
2.1 Choose the size neces-
sary for the application.
Orange colour code: 2.5 l,
yellow colour code: 1.5 l
2.2 Attach the jar to the carry-
ing handle (portable
version only) or to the
optional clampholder on
the trolley (see step 5.4).
3.1 Prepare and insert corre-
sponding liner size
according to instructions
delivered with liners. The
disposable liner has an
integrated overflow pro-
tection filter. No action is
necessary to activate it.
4.1 Connect tubing from the
jar (A) to the Safety Set
(B).
5.1 Connect patient tubing to
the lid of the liner using
either the coloured angle
piece or the patient port
directly (depending on
tubing size).
1.1 Connect the equipotential
conductor at the back of
the Basic 30 with the bus
bar as advised by your
technician and according
to inhouse directives. The
respective cable is not
supplied from Medela.