CALIBRATION PATH
Your Pro Plus scale has been calibrated at the factory and does not require
re-calibration prior to use.
Calibration is performed using kg or lb, according to the units of measure used upon
entering into the calibration path.
01-Cal With One Point process below:
Operator Action
1. Make sure the scale is turned off, then press
and hold the HOLD/RELEASE button while you
press the On/OFF button.
2 Input the weight , then press ENTER button.
3. Make sure there is nothing on the platform, then
press ENTER button.
4. Wait for about 1-3 second, Zero calibration
finished.
5. Put the weight on the platform, then press
ENTER button.
6. It will calibrate after 1-3 seconds.
7. If success, display shows Factor OK!. If failed,
display Factor Fail.
8. If successful the process will clear the weight.
9. The scale will reboot after wait for about 1-2
seconds.
10. Please wait until the scale resumes normal
operation. Remove the weight from the platform.
Enter load weight:
450.0LB( or 200.0KG)
Enter load weight:
450.0LB( or 200.0KG)
Zero calibration
Please clear the scale
Zero calibration
Please wait
Weight calibration
Put: XXX.X
Weight calibration
Please wait
Factor Update
Factor OK! or Factor Fail!
Rebooting...
Please wait
Health o Meter
Pro-Plus
Health o Meter
Pro-Plus
24
Display