If S8 is on, then the optocoupler output functions as a pulse output. For pulse values, refer to table.
If S8 is off, then the optocoupler output functions as a contact output and signals the forwards/reverse
direction of flow.
If working with the version without a display, you have the option of setting the measurement range in
l/min or in US gal/min. The switch position for the relevant measurement range can be found in the fol-
lowing tables. The factory presetting is: all switches to ON.
Measurement range setting in l/min (S4 on)
D184B112U04
FXL5000 (Miniflow)
Damping
S5
On
=
5 seconds
S5
Off
=
10 seconds
Measurement range
and pulse output
S4
On
=
Litre
S4
Off
=
US gallon
Current output
S7
On
=
4−20 mA
S7
Off
=
0−20 mA
Pulse output/contact output (F/R)
S4 on and
S8 on
=
S4 off and
S8 on
=
S8 off
=
DN
Stage 1
S1
S2
S3
Off
Off
Off
10
2.5 l/min
15
5
l/min
25
15
l/min
50
60
l/min
DN
Stage 1
S1
S2
S3
On
Off
Off
10
12.5 l/min
15
25.0 l/min
25
75.0 l/min
50
300.0 l/min
1 pulse per 1 litre
1 pulse per gallon
Contact output
Signalling in direction of flow
Forwards/reverse (F/R)
Stage 2
S1
S2
S3
S1
Off
Off
On
Off
5
l/min
7.5 l/min
10
l/min
15.0 l/min
30
l/min
45.0 l/min
120
l/min
180.0 l/min
Stage 2
S1
S2
S3
S1
On
Off
On
On
15
l/min
25
30
l/min
50
90
l/min
150
360
l/min
600
Stage 3
Stage 4
S2
S3
S1
S2
On
Off
Off
On
10
l/min
20
l/min
60
l/min
240
l/min
Stage 3
Stage 4
S2
S3
S1
S2
On
Off
On
On
l/min
50
l/min
l/min
100
l/min
l/min
300
l/min
l/min
1200
l/min
6 Operation
S3
On
S3
On
GB 17