Assembling the sensor
4
3
Fig. 2: Electrolyte filling
1
Membrane cap
2
Electrolyte filling height
3
Pipette
4
Vent hole
NOTICE!
Perform the following work over a
sink.
1.
Remove the end cap fully from the
nozzle and cut open the nozzle at
the marking to open up the nozzle
duct
2.
Remove the membrane protective
cap and unscrew the membrane
cap from the electrode shaft
3.
Fill the membrane cap up to the
bottom thread
4.
Wait approx. 1 minute until the elec‐
trolyte has completely covered the
sensor
5.
Place the electrode shaft vertically
onto the filled membrane cap
6.
Turn the electrode shaft in such a
way that the vent hole points
upwards
7.
Do not close the vent hole beneath
the hose seal using your fingers
46
8.
Slowly turn the membrane cap by
hand up to the stop. Excess electro‐
lyte will escape through the vent
hole during screwing together
9.
Rinse any leaked electrolyte from
the sensor and your fingers under
running water