E. MANUAL OF THE MONITEUR (Option)
Your UV equipment is fitted with a UV monitor and a UV-C radiation measurement cell.
This monitor will indicate the gradual reduction in the intensity of the lamps throughout their lifetime but
may also provide a warning of dirt being deposited in the quartz sleeve or on the UV-C radiation
measurement cell.
OPERATION
1. Switch on the lighted switch on the front of the electrical unit..
2. The UV C lamps will heat up to their maximum output level in 2 to 5 minutes (depending on the
temperature of the liquid being treated).
3. You now need to calibrate the cell in its liquid environment:
1. Open the front of the electrical unit.
2. Take a small screwdriver.
3. Turn gently the small screw at the back of the monitor (next to a green LED) clockwise
until the LED light green.
Calibration should result in a green fixed LED on the front and back face
UV_NM-GB.doc
Marque, Modèles et Brevets déposés - Produits exclusifs
UV MONITOR
UV INTENSITY
< 50 %
Main
Alarm
CALIBRATION of the UV-C ray measurement cell
(Do it at each change of lamp and cleaning of sensor)
of the UV Monitor.
50...70 %
> 70 %
Pre
OK
alarm
Copyright BIO-UV - 29/09/2008