Calibration; Scale Calibration - Health O Meter 349KLX User Instructions

Ocultar thumbs Ver también para 349KLX:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

CALIBRATION

Calibration
This scale has been factory calibrated and does not require calibration prior to use. If
required, the scale can be calibrated. Calibration of the 349KLX scale is performed using
certified weights. 50 / 100 / 200 / 400 lbs are employed as the calibration standards. The scale
can only be calibrated in pounds (lb).
Caution: To avoid injury, use extreme care when stacking weights on platform.
Note: Only weights certified and traceable to national standards should be used for
calibration procedures.

Scale Calibration

Press and hold the lb./kg and ZERO buttons at the same time, then
press the ON button to power on the scale. "349" will briefly appear on
the display, followed by "CAL" and "UXX". (Two alphanumeric characters
will appear next to "U".) Release all three buttons. Several alphanumeric
characters will appear on the display (1-4 characters can be displayed
during calibration).
Press the ZERO button, and "C-50" will flash on the display, followed by
several alphanumeric characters.
Place 50 lbs on the scale. Several alphanumeric characters will appear.
When calibration characters stabilize, press the ZERO button. "C-100"
will appear on the display, followed by several alphanumeric characters.
Place 100 lbs on the scale. Several alphanumeric characters will appear.
When calibration characters stabilize, press the ZERO button. "C-200"
will appear on the display, followed by several alphanumeric characters.
Place 200 lbs on the scale. Several alphanumeric characters will appear.
When calibration characters stabilize, press the ZERO button. "C-400"
will appear on the display, followed by several alphanumeric characters.
Place 400 lbs on the scale. Several alphanumeric characters will appear.
When calibration characters stabilize, press the ZERO button. "END" will
appear on the display, followed by "OFF". The scale will automatically
shut off.
ACTION
16
DISPLAY SHOWS
"349" followed by
"UXX" then
"XXXX"
"C-50" then "XXXX"
"XXXX" followed
by "C-100" then
"XXXX"
"XXXX" followed
by "C-200" then
"XXXX"
"XXXX" followed
by "C-400" then
"XXXX"
"XXXX" followed
by "END" then
"OFF"

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido