• Pressing the key for one second will lead to the peak measurement mode,
"PEAK" and "MANU" will appear on top right position of LCD display. Press
and hold the key for one second once again, "PEAK" will disappear from the
top right position of the LCD interface, and "AUTO" will appear in the middle
of LCD, and the mode will return to auto measurement mode.
• Press and hold max/min value inquiry key (5) for a short time, "MAX" will
appear on top position of LCD interface, switching the device to the max/min
value inquiry mode. In this mode, each short press of the key causes the
mode to change from MAX to MIN or from MIN to MAX; press and hold the
"MAX/MIN" key (5) for one second, "MAX/MIN" on top position of the LCD
interface will disappear, and the mode will exit.
• Press and hold the data hold/zero calibration key (7) of the KPS-LX30LED
for a short time to enter the data hold mode, which will be indicated by the
apparition of "HOLD" on the top left position of the LCD interface; press and
hold the key for a short time once again, "HOLD" will disappear from the top
left position of the LCD interface, and the device will exit the data hold mode.
• Press and hold the key for a short time once again, "HOLD" will disappear
from the top left position of the LCD interface, and the device will exit the
data hold mode.
• In any mode with sensor covered, press and hold data hold/zero calibration
key (7) for one second, "ADJ" will appear on the LCD interface, and the
device will enter the zero calibration mode: several seconds later "ADJ" will
disappear from the LCD interface, and the device will exit the data hold
mode and return to auto measurement mode.
• The reference position of light source test is at top of sphere surface under
light.
• In various measurement modes, analog bar in the middle of the LCD screen
will change with figures of Nixie tubes.
• When reading, existing data can be locked by pressing and holding "HOLD/
ZERO" key for a short time.