Guidance and manufacturer's declaration-electromagnetic immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration-electromagnetic immunity
The Oxystart
Pulse Oximeter is intended for use in the electromagnetic environment specified below. The
®
customer or the user of the Oxystart
Immunity test
IEC60601 test
level
Radiated RF
3V/m
ICE 61000-4-3
80MHz to
2.5GHz
Pulse Oximeter should ensure that it is used in such an environment.
®
Compliance
Electromagnetic environment - guidance
level
3 V/m
Portable and mobile RF communication equip-
ment should be used no closer to any part of
the Oxystart
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in meters (m).
Pulse Oximeter,including cables,
®
80 MHz to 800 MHz
3
5 .
d
=
P
E
1
7
800 MHz to 2,5 GHz
d
=
P
E
1
Oxystart® - EN