5. FUSE REPLACEMENT
a)
CAUTION:
Before attempting to change fuses turn the
master power switch S1 "OFF" (O) and disconnect amalgamator
unit from AC outlet (Mains).
b) Switch off master switch S1 and unplug the line cord. Pry the fuse
holder open with a small screwdriver as shown in Figure 5b-1.
Remove the fuses from the holder as shown in Figure 5b-2 and check
to see which fuse is blown. The thin wire in the center of the fuse will
be broken or melted away if it is a defective fuse. Replace the fuse in
the fuse holder and firmly push the fuse holder back into the power
entry block.
c) Replace fuses with same type and value:
115/120V: T 0.63A/250V, part no. 930847
230V: T 0.315A/250V, part no. 930846
6. CLEANING, CARE AND MAINTENANCE
CAUTION: Before cleaning unit turn master power switch
"OFF" (O) and disconnect unit from AC outlet (Mains)!
CAUTION: Do not allow cleaning solution to run into unit
through the power switch, fuse door, or capsule arm opening. this
may damage unit and/or cause electrical shock.
CAUTION: Do not use a bleach solution to clean case, as this
will yellow the surface finish.
Clean the OptiMix case using a soft cloth and Cavicide cleaning solution.
Follow recommendations on Cavicide container.
•
Use care while wiping windows to avoid scratches.
•
Your OptiMix has been lubricated at the factory and will not
require lubrication for the life of the machine.
•
Calibration of unit is not required as on some amalgamators.
11
7. SPECIFICATIONS AND EQUIPMENT CLASS
a)
AC supply connection:
b)
Power input:
c)
Equipment class:
d)
ISO 7488
Amalgamator Class:
e)
Protection from
ingress of liquids:
f)
Operational Environment:
• Ambient temperature
• Relative humidity
• Atmospheric pressure
g)
IEC 1010 Installation category:
h)
IEC 1010 Pollution Degree:
i)
Transport and storage:
• Ambient temperature
• Relative humidity
• Atmospheric pressure
j)
Operation:
k)
Fuses (2 per unit):
l)
Unit overall dimensions:
• Height
• Length
• Width
m)
Unit weight:
n)
Additional symbols:
•
Attention, consult
accompanying
documents
•
Alternating Current
115/120V~/0.5 A, 50/60 Hz
230V~/0.25A, 50/60 Hz
Note: Mains supply voltage fluctuations are not to exceed
10% of nominal rated supply voltage
100 VA, max.
I
Type 2 Variable Frequency
None
59°F to 104°F (15° C to 40°C)
10% to 100%, including condensation
.5 atm to 1.0 atm (500 hPa to 1060 hPa)
Table-top equipment for indoor use only
Equipment not suitable for use in the presence of
flammable anaesthetic mixture with air or nitrous oxide
II
2
-40° F to 158° F (-40° C to 70° C)
10% to 100%, including condensation
.5 atm to 1.0 atm (500 hPa to 1060 hPa)
Continuous operation with intermittent loading,
duty cycle of 0.5 minutes on/2.0 minutes off
115/120V: T 0.63A/250V, part no. 930847
230V: T 0.315A/250V, part no. 930846
5"
10"
7 5/8"
8.25 lbs.
• OFF
• Protective Earth
• ON