6. DELETING THE MEMORY
With a flashing "
skip. To delete all the results, press UP and then a flashing "dEL" and "
the display, press UP again to delete all the memory records.
7. SETTING THE REMINDER ALARM
Your meter has four reminder alarms. The meter will display "OFF", "1st" and "
If you don't want to set an alarm, press REPEAT to select "OFF" to skip this step.
If you want to set an alarm, press REPEAT to select "On".
With the hour/minute flashing in sequence, press UP or DOWN to select the desired
hour/minute. Press REPEAT and go to the next alarm setting.
NOTICE:
When the alarm beeping: Press M to silence it; Or press and Hold M to switch it off; Or it will
beep for 2 minutes then switch itself off.
NOTE:
• These parameters can ONLY be changed in the setting mode.
• If the meter is idle for 3 minutes during the setting mode, it will switch off automatically.

4. THE TWO MEASURING MODES

The meter provides you with two modes for measuring, General and CTL. You can switch
between each mode by:
1. Start with the meter switched off. Insert a test strip to turn on the meter. The screen will
display a flashing " " and "
2. Press M to switch between to CTL mode.

5. BEFORE TESTING

5.1. CONTROL SOLUTION TESTING

Our Control Solution contains a known amount of glucose that reacts with test strips and is
used to ensure your meter and test strips are working together correctly.
Do a control solution test when:
• you first receive the meter,
• at least once a week to routinely check the meter and test strips,
• you begin using a new vial of test strips,
• you suspect the meter or test strips are not working properly,
• your blood glucose test results are not consistent with how you feel, or if you think the
results are not accurate,
32
" on the display, press REPEAT to keep the results in memory to