Potential Adverse Effects:
Cataracts.
Male sterility.
Enhanced drug activity.
Thermal stress.
4.2
Parameters
Ultrasound Frequency, expressed in MHz, is the frequency of the ultrasound waves. The ultrasound
frequency determines the penetration depth, which has the largest value at 1 MHz. The ultrasound
frequency can be set at 1 MHz or 3 MHz.
Duty Cycle, expressed in %, defines the ratio of the pulse duration to the pulse repetition time.
Ultrasound can be applied in pulsed or in continuous mode.
Effective Radiation Area (ERA) expressed in cm², defines the cross-sectional area of the ultrasound
beam (See technical specifications for details). The Effective Radiation Area is fixed and defined by the
size of the ultrasound applicator.
Ultrasound Power is the ultrasound output expressed in W. The ultrasound output display can be
toggled between W and W/cm². In pulsed mode the power during the pulse is displayed. The time
averaged power can be obtained by multiplying this value with the Duty Cycle.
Ultrasound Amplitude, expressed in W/cm², is the quotient of Ultrasound Power and Effective Radiation
Area. The ultrasound output display can be toggled between W and W/cm². In pulsed mode the
Amplitude during the pulse is displayed. The time averaged Amplitude can be obtained by multiplying this
value with the Duty Cycle.
5 Package Contents
1631901
1631902
1631903
1631904
5.1
Standard Accessories for 1631.901, 902 and 903
3444357
1631810
3442929
1631750
(1) = The Sonopuls is delivered with 1 bottle of contact-gel. The article number 3442929 however represents a box of
12 bottles.
5.2
Standard Accessories for 1631.904
3444357
0169842
3442941
3442942
1631750
1629750
1929750
5.3
Optional accessories for Ultrasound / StatUS™ Therapy
3442929
40
3442930
3442931
Sonopuls 190 incl. multi-frequency ultrasound head of 5 cm2
Sonopuls 190 incl. multi-frequency ultrasound head of 0,8 cm2
Sonopuls 190 incl. 2 multi-frequency ultrasound heads (5 cm2 and 0,8 cm2)
Sonopuls 190 incl. StatUS™ Pack 100
Mains connecting cable
Head holder for multi-frequency treatment head
Contact-gel, bottle of 250 ml
User manual Sonopuls 190 (CD-Rom)
Mains connecting cable
Head holder StatUS™ applicator
Gelpad StatUS™ (240x)
Fixation ring gelpad StatUS™ (3x)
User manual Sonopuls 190 (CD-Rom)
User manual StatUS™ Pack 100 (CD-Rom)
Information booklet StatUS™ Pack 100
Contact-gel, bottle 250 ml, box of 12
Contact-gel, bottle 850 ml, box of 12
Contact-gel, canister of 5 l