Before Reprocessing; Patient-Specific Third-Party Accessories And Consumables; Validated Reprocessing Procedures; Surface Disinfectants - Dräger PressurePod Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 32
English / English
USEnglish / English US

Before reprocessing

Observe before disassembly
1
Screw sealing caps onto the pressure measuring
ports.
2
Remove the USB plug of the PressurePod from
the USB port of the connected compatible de-
vice.
Patient-specific third-party accessories
and consumables
The patient-specific third-party accessories and con-
sumables must be removed from the device and, if
necessary, disassembled.
Reusable products:
 If the reusable product has its own instructions
for use, perform reprocessing in accordance with
the separate instructions for use.
Disposable products:
 Dispose of the disposable products.

Validated reprocessing procedures

Overview of the reprocessing procedures of the
components
Components
Surface disinfection with
cleaning
Device surface
Yes
18
Surface disinfection with cleaning
Surface
Manufac-
Concen-
disinfec-
turer
tration
tant
Dismozon
BODE
1.6 %
pur/plus
Chemie
Oxycide
Ecolab
2.3 %
USA
Prerequisites:
The surface disinfectant has been prepared in
accordance with the manufacturer's instructions.
The manufacturer's instructions, e.g., regarding
shelf life or application conditions, are observed.
An uncontaminated, lint-free cloth soaked in sur-
face disinfectant is used for the cleaning surface
disinfection.
CAUTION
Risk due to penetrating liquid
Penetrating liquid may cause the following:
Damage to the device
Device malfunctions
Ensure that no liquid penetrates the device.
Cleaning
1
Wipe off obvious soiling with a disposable cloth
soaked in surface disinfectant. Dispose of the
cloth.
2
Wipe all surfaces. After that, there must no lon-
ger be any soiling visible.
Surface disinfection
3
Wipe cleaned surfaces again to visibly wet all
surfaces to be disinfected with surface disinfec-
tant.
4
Wait for the surface disinfectant contact time.
5
At the end of the contact time, moisten a new, un-
contaminated and lint-free cloth with water (at
Contact
least drinking water quality).
time
6
Wipe all surfaces until no remains of the surface
disinfectant, such as foam residues or streaks,
15 min
are visible.
7
Wait until the surfaces are dry.
5 min
8
Check the surfaces for visible damage and, if
necessary, replace the product.

Surface disinfectants

In addition to the surface disinfectants mentioned in
the "Validated reprocessing procedures" section,
other surface disinfectants are listed on the following
web page:www.draeger.com/disinfectants

Storage and transport

After reprocessing, there are no special require-
ments for storage and transport of the product. How-
ever, the following must be observed:
Store dry and free of dust
Avoid recontamination and damage during trans-
port
All further information on storage and transport in-
cluded in the accompanying documents must be ob-
served.
Instructions for use PressurePod

Publicidad

Tabla de contenido
loading

Tabla de contenido