Guidance and manufacturer's declaration – electromagnetic immunity
+
The ClearReader
is intended for use in the electromagnetic environment specified below. When
usingthe ClearReader
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
interruptions and
English
+
, make sure the ClearReader
IEC 60601
test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output
lines
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
<5 % UT
+
is being used in the environment described below.
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output
lines
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
<5 % UT
ClearReader
Electromagnetic
environment –
guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30 %.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
+