Warnings; Error Reports - B.Braun Aesculap GN200 Instrucciones De Manejo/Descripción Técnica

Ocultar thumbs Ver también para Aesculap GN200:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 62
Aesculap
Lektrafuse HF generator GN200
7.2

Warnings

A warning is indicated as follows:
Individual tone (different from a failure, an error or normal HF opera-
tion)
Warning message in the display 5 on the front side of the HF generator
Warning message
Generator Cooling
Please Wait
Generator Too Cold
Please Wait
7.3

Error reports

An error is indicated as follows:
Unambiguous tone sequence (different from a failure or normal HF
operation)
Failure message in the display 5 on the front side of the HF generator
Error code
Error report
E001
ERROR E001
Instrument HF activation button error
Release Instrument RF-ON Button
Remove and reattach instrument
E002
ERROR E002
Instrument Error
Remove and reattach OR
Remove and replace instrument
E003
ERROR E003
Footswitch Error
Release footswitch
Remove and reattach footswitch
16
®
Significance
generator temperature too high
Generator must first warm up
Remedy
Check the airflow around the generator
Do not switch off the generator whilst it is warming up
Significance
Instrument "HF activation" button is
pressed while the instrument is con-
nected to the HF generator
HF generator cannot find instrument
- or -
Instrument is defective and must be
replaced
Foot switch is pressed while the foot
switch is connected to the HF generator
Remedy
Release the "HF activation" button
Remove instrument and reconnect it
Remove instrument and reconnect it
- or -
Replace instrument
Release foot switch
Remove foot switch and reconnect it

Publicidad

Tabla de contenido
loading

Tabla de contenido