20
*THESE CHECKS SHOULD INCLUDE:
1. SPREADER BAR: Check the spreader bar for freedom of rotation
and swing. Check for wear on the central pivot. Check for firm
attachment to the boom.
2. BOOM: Check the attachment of the boom to the mast carriage.
Make sure there is only minimal movement.
3. MAST: Check it is firmly connected to the chassis.
4. MOTOR GEARBOX: This should require no maintenance other than
checking for correct operation and listening for unusual noise.
5. POWER PACK: Check the function of the emergency stop button
and emergency down/up.
6. BATTERIES: The batteries are housed in the power pack and should
not require maintenance other than the regular charging as
detailed in the charging instructions. Check that the connections
remain clean. Confirm the battery bracket is secure and undamaged.
7. LEG ADJUSTMENT: Check that the legs operate in both full
extensions (inward/outward).
8. CASTORS: Check all castors for firm attachment to the legs.
Check for free rotation of the castor and the wheels.
9. CLEANING: Wipe down with a clean cloth, using soap and water
and/or any hard surface disinfectant. Harsh chemical cleaners or
abrasives should be avoided as these may damage the surface
finish of the lift. Avoid excessive wetting, especially electrical
areas and the opening to the base of the hoist. We recommend
that the cloth is soaked in the solution, squeeze out excess fluid
and wipe hoist surface.
10. LOAD TEST: The load test should be carried out in accordance
with the manufacturer's test procedures (see Service Manual).
It is strongly recommended the testing is carried out by an
authorised service dealer.
11. BASE AND WHEELS: Ensure base is even and level
(all four wheels are on the floor).
12. SLINGS: Check for wear and fraying.
13. LUBRICATION: Oil pivot joints, including mast and boom
connections, pedal assembly, spreader bar joint (only if required).
14. HAND SET: Ensure plugged fully into controller.
294000.10017 Rev. B
E n g l i s h
E n g l i s h
*THESE CHECKS SHOULD INCLUDE:


15. QUICK RELEASE PIN: Ensure the pin is securely fitted
before lifting, by pressing the barbed end.
16. FASTENERS: Check fasteners for tightness.


17. LIFTING STRAP: Your lift strap should be checked at service
intervals for any sign of wear (ideally every six months). In order


to view the strap it is recommended to remove the cap at the top
of the mast, cycle the lift in both directions whilst observing the

condition of the strap. If you see any wear (fraying, creasing or
similar) it is recommended an authorized, Sunrise Medical Dealer


or qualified, competent service engineer replaces the strap
IMMEDIATELY. For details regarding strap maintenance and
replacement please request our detailed service guide.

(See below for details.)