Air filter
Take the air filter out of the compressor and inspect it at regu-
lar intervals (after every 10 applications). If it is discoloured
brown or grey, or if it is damp or clogged, replace it.
The filter must be replaced after 200 operating hours (ap-
prox. 1 year) at the latest [see: Replacing the air filter,
page 31].
The air filter cannot be cleaned and then reused!
4.2 Multiple patient use
Cleaning and disinfection cycles
Nebuliser with
PARI SINUS tubing
system and air filter
Compressor housing
Nebuliser
The nebuliser is not suitable for use by more than one patient.
Use a separate nebuliser (including connection tubing) for each
patient.
Air filter
All air filters must only be used for one patient. Replace the air
filter before every patient change [see: Replacing the air filter,
page 31].
Disinfecting the compressor housing
For disinfection, use a standard, alcohol-based disinfectant
(e.g., isopropanol). Regarding the use and quantity of the disin-
fectant, it is imperative to follow the information for use of the
agent used.
To disinfect the compressor, proceed as follows:
• If the compressor is visibly dirty, clean the compressor be-
fore disinfecting it [see: Cleaning the compressor housing,
page 29].
– 30 –
Replace before each patient change
Disinfect before each patient change