On following uses, it will display the first file in the system last used before switching the device
off or the last file used before switching into standby mode.
Standby Mode
After the device has not been used for 10 minutes, the device will automatically enter the
standby mode to conserve battery power. The display will switch off and the green LED on the
power key will flash. To exit the standby mode, press any key or the foot pedal: the device will
switch on and resume functionality from the last screen displayed prior to entry in standby.
After 30 minutes in standby mode, the device will switch off to conserve battery power. It can
be switched on again by pressing the POWER key.
Switch-off
Press the POWER key to switch off the device. The battery LED (2) (See Fig. 4) will not light
up unless the battery is charging.
6.5.2. Calibration
Calibrating CAL (8) (See Fig. 4) the micromotor automatically adjusts the torque measurement
to ensure torque accuracy each time the micromotor has been changed, the contra-angle is
replaced, autoclaved and/or lubricated.
Connect the WaveOne
Hold down the CAL (8) key (See Fig. 4).
The micromotor will turn from the minimum to the maximum speed to measure the contra-
angle inertia automatically.
WARNINGS
Never press the calibration button during treatment.
During calibration, the micromotor will vary its speed from the minimum value to the
maximum value. Do not insert any file. The calibration process can be aborted by pressing
any key or foot pedal.
Calibrate every time the contra-angle is lubricated or replaced after sterilizing, or at least
once a week.
Do not use any other contra-angle or other reduction rate than the original one.
Refer to the separate operating instructions of WaveOne
before lubricating it.
www.enversaltan.com
(MAILLEFER 6:1) contra-angle to the micromotor (without a file).
™
www.enversaltan.com
(MAILLEFER 6:1) contra-angle
™
- 17 /141 -