ORGANIGRAMA PARA LA SOLUCIÓN DE PROBLEMAS
ORGANIGRAMA PARA LA SOLUCIÓN DE PROBLEMAS
Figura D.1 Organigrama para la solución de problemas
Instrument
operation mode
Check
Check
Printer
Reagents
Level
Connected
check
Yes
Yes
Switched
Expiration
on
date
Yes
Yes
Paper
Waste Level
feed
check
OK
See Heading A.10,
See Replacing
See
Reagents
Replacing
Printer Configuration
the Waste
Container
Incorrect
results
All
RBC, PLT, HCT
parameters
Reproducibility
Reproducibility
Yes
Calibrated
Calibrated
OK
See Chapter 6,
See Chapter 6,
Reviewing Results
Reviewing Results
D-2
Instrument
startup
No
Main supply
Connect
connected
Yes
No
Press ON/OFF
Switched on
switch
Yes
Failed
OK
Startup
See Startup
passed
Yes
No
Temperature
See Startup
reached
Yes
Ok
Calibration
verification
(QC)
Poor
See Chapter 7,
Calibration
HGB
Reproducibility
Yes
Yes
Calibrated
OK
OK
See Chapter 6,
See Chapter 6,
Reviewing Results
Reviewing Results
PN 4237633DC
Analysis
cycle
Instrument
Sampling
alarms
Bent
Technical
Probe
operations
OK
OK
error
message
No
Blood
Temperature
aspiration
OK
OK
No
Probe
Reagents
motion
OK
OK
Printout/
transmission
Contact a
Beckman Coulter
representative.
See Heading 8.6,
System
Troubleshooting
Procedures
WBC
DIFF
Reproducibility
Reproducibility
Yes
Yes
Calibrated
OK
See Chapter 6,
Reviewing Results
See Heading 6.3,
Flags Generated
by the Instrument
Dilution
No
Traverse
motion
Yes
No
Sample
distribution
Yes
OK
No
Drain
Yes
Rinse
See Heading 8.6,
System
Troubleshooting
Ok
Procedures
Flags
Instrument
Interpretive
flags
messages
WBC
WBC
OK
OK
RBC
RBC
OK
OK
Plt
Plt
OK
OK
Hgb
Hgb
OK
OK
OK
OK
DIFF
See Heading
DIFF
6.4,
Interpretive
Messages
Flags
WBC
OK
RBC
OK
Plt
OK
Hgb
OK
DIFF
See Chapter
6, Reviewing
Results