7. Maintenance
After each patient maintain the product as follows.
7-1 Cleaning
1) Remove dirt and debris from the product. Do not use a wire brush.
2) Wipe clean with alcohol-immersed cotton swab or cloth.
This icon denotes that the product can be washed via Thermo Disinfector.
Refer to the Thermo-Disinfector manual.
CAUTION
· After washing with Thermo-Disinfector and prior to lubrication, dry the product until all internal moisture is thoroughly removed.
Thermo-Disinfector moisture remaining inside the product could reduce the effect of lubrication and could cause corrosion inside of the
product.
· To clean the product never use any solvent such as benzine or thinner.
6
7-2 Lubrication
NSK PANA SPRAY Plus
Apply NSK PANA SPRAY Plus every time after each use and/or before autoclaving.
1) Remove the motor from the hose.
2) Mount the arrow-head spray nozzle tip into the spray can port. (Fig. 6)
3) Insert the Spray Nozzle into the Drive Air Port of the motor. Hold the motor and spray for approximately
2-3 seconds. Apply lubricant until it expels from the exhaust air part for at least 2 seconds. (Fig. 7)
CAUTION
· When applying spray be sure to hold the motor firmly to prevent the motor from slipping out of the
hand due to the spray pressure.
· Hold the spray can upright.
Arrow-head Spray Nozzle
Motor
PANA SPRAY Plus
Fig. 6
Exhaust Air Part
Drive Air Port
Fig. 7