Scope of Supply:
1 pc. Continuity Tester VT CTL
2 pcs. AAA Batteries
1 pc. User Manual
2.0 Transport and Storage
Please keep the original packaging for later transport, e.g. for calibra-
tion. Any transport damage due to faulty packaging will be excluded
from warranty claims.
Instruments must be stored in dry and closed areas.
In the case of an instrument being transported in extreme tempera-
tures, a recovery time of minimum 2 hours is required prior to instru-
ment operation.
3.0 Safety
Continuity Tester have been constructed and verified in compli-
ance with the latest safety standards for Test Instruments IEC/
EN 61010-1 and IEC/EN 61010-031 and have left the factory in
safe and perfect conditions.
In order to avoid electrical shock, the valid safety and VDE
regulations regarding excessive contact voltages must receive
utmost attention, when working with voltages exceeding 120V
(60V) DC or 50V (25V)rms AC. The values in brackets are valid
for limited ranges (as for example medicine and agriculture).
Prior to usage ensure perfect instrument function.
The safety can no longer be insured if the instrument:
• shows obvious damage
• does not carry out the desired measurements
• has been stored for too long under unfavourable conditions
• has been subjected to mechanical stress during transport.
The UUT must be voltage-free. Verify this condition, if unsure,
by using a voltage tester equipped with a display.
9