CAUTION
• To avoid product failure, do not use a sterilizer that exceeds a cycle temperature of 138°C, including the dry cycle. In some
sterilizers, the chamber temperature may exceed 138°C. Contact the sterilizer manufacturer for detailed information about
cycle temperatures.
• Clean the light probe prior to autoclaving. If debris remains on the internal surface it can get stuck and cause product failure.
• Do not autoclave the product with other instruments even when it is in a pouch. This is to prevent possible discoloration and
damage to the product from chemical residue on other instruments.
• Do not heat or cool the product too quickly. Rapid change in temperature could cause damage to the product.
• Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. The air should be free
from dust, salt and sulphur.
• Do not touch the product immediately after autoclaving as it will be very hot and must remain in a sterile condition.
• Autoclave sterilization is recommended for the product. The validity of other sterilization methods is not confirmed.
NOTICE
• NSK recommends Class B sterilizers as stated in EN 13060.