Mestra R-080110 Manual Del Usuario página 14

Horno de circonio
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

English
If we want to store the programs of the example graphics (Fig. 10) and carry out the second program, the
parameters should be the following:
Step
Parameter
C1
Step 1
t1
C2
Step 2
t2
C3
Step 3
t3
C4
Step 4
t4
C5
Step 5
t5
C6
Step 6
t6
C7
Step 7
t7
C8
Step 8
t8
C9
Step 9
t9
C10
Step 10
t10
C11
Step 11
t11
14
Value
Meaning
0 ºC
Initial temperature
-8
Step in which the furnace will start working
0 ºC
Initial temperature
30 min.
Time selected to reach the next temperature
400 ºC
First set-up temperature
15 min.
Time selected to reach the next temperature (maintenance)
400 ºC
First set-up temperature (maintenance)
30 min.
Time selected to reach the next temperature
1000 ºC
Second set-up temperature
15 min.
Time selected to reach the next temperature (maintenance)
1000 ºC
Second set-up temperature (maintenance)
15 min.
Time selected to reach the next temperature
600 ºC
Third set-up temperature
-121
End
0 ºC
Initial temperature
15 min.
Time selected to reach the next temperature
500 ºC
First set-up temperature
60 min.
Time selected to reach the next temperature
1500 ºC
Second set-up temperature
20 min.
Time selected to reach the next temperature (maintenance)
1500 ºC
Second set-up temperature (maintenance)
-121
End
Fig 10

Publicidad

Tabla de contenido
loading

Tabla de contenido