SPECIFICATIONS
Model No.: TD-4251
Dimension & Weight: 86.9mm(L) x 51mm(W) x 15.8mm(H), 42.2g
Power source: One CR2032
Memory: 450 measurement results with date and time
Auto electrode inserting detection
Auto sample loading detection
Auto reaction time count-down
Auto turn-off after 3 minutes without action
Operating condition: 50°F ~104°F(10°C~40°C), below 85% R.H.(noncondensing)
Meter storage/Transportation condition: -4°F~140°F(-20°C~60°C), below 95% R.H.
Measurement Units: Fix mg/dL
Measurement Range: 20~600mg/dL (1.1~33.3mmol/L)
The device has been certified to meet the electrical and safety requirements of:
IEC 60601-1, EN 60601-1, IEC 61010-1, EN 61010-1, EN 61010-2-101, EN
60601-1-2, EN 61326.
Performance characteristics
Accuracy
Within ±15 mg/dL (0.83 mmol/L) at glucose concentration < 75 mg/dL
(4.2 mmol/L) and within ±20% at glucose concentration ≥ 75 mg/dL (4.2
mmol/L).
Precision
CVs (%) of intermediate precision and repeatability are less than 5%.
The product has been tested to meet the requirements of ISO 15197.
51