Example: The blank value of a chloride titration is defined with the support of an extra method. The result in ml is
thereby automatically written into global memory M01 by using the name "Blanc value" (Fig. 128). The blank
value is then automatically deducted from the titrant consumption within the chloride method.
Fig. 128
The menu for the global memory can always be accessed by pressing <Shift> or via system settings. The name
or values can be changed by using <EDIT> and have the methods shown that are used in the global memories
(Fig. 129).
Fig. 129

Formula Editor

4.6.4
The formula editor is intended to complement the existing standard formulas. The standard formulas are loaded
and can then be modified. The original standard formula itself is never changed.
4.6.4.1 Start and Work with the Formula Editor
With <EDIT> you are going to «Edit method», «New method» or «Default methods» and select then «Result»
(Fig. 130).