235
4.7.2.1 Initial waiting time (Dead-Stop titration) / Extraction time (KF)
With Dead-Stop titration, the «Initial waiting time» passes at the beginning of titration. In KF titration, the Initial
waiting time = the «extraction time». The extraction time ends after the sample is supplied. The initial
waiting/extraction time can be specified between 0 and 999 seconds (Fig. 189).
Fig. 189
4.7.2.2 Conditioning (only KF)
«Conditioning» is activated for every KF method. It can be shut off via a PC for external control (Fig. 190)
Fig. 190
4.7.2.3 Fixed delay time
The «fixed delay time» is the waiting time between the linear titration steps at the end of the titration until the
Endpoint. The fixed delay time can be set between 0 and 999 seconds (Fig. 191).
Fig. 191