Fig. 148
If normal or fast drift was selected before, the values will be defaulted for user-defined drift. In the present case,
for instance, 20 mV for normal drift: (Fig. 149).
Fig. 149
Drift-controlled acceptance of the measured value is used in most applications.
However, there are applications in which the setting of a fixed holding time for measured value acceptance
following the titration step is recommendable. Examples hereof include titrations in non-aqueous media. In the
case of Dead-Stop titration no holding time other than the fixed one can be selected. The fixed delay time can be
set between 0 and 999 seconds (Fig. 150).
Fig. 150