Maintenance
Maintenance ....................................17
Mantenimiento .................................18
維護 .................................................18
Техническое обслуживание ..........18
Maintenance Interval
EN
Every 300 uses or 3 months
Maintenance Procedure
• Lubricate internal Check Valve by removing pumping piston and internal
Check Valve. Squirt 2 ml of oil into the threaded end of the Check Valve
and reassemble. When the pumping piston is pumped the oil will be
pumped through the internal Check Valve.
• Lubricate external Check Valve by removing external Check Valve. Squirt
2 ml of oil into the threaded end of the Check Valve and reassemble.
When the pumping piston is pumped the oil will be pumped through the
external Check Valve.
• Lubricate the Ralston Quick-test™ fittings by squirting 2 ml of oil inside
the connection.
• Lubricate the pumping piston and the fine adjust piston O-rings with
silicone lubricant.
Intervalle d'entretien
FR
Tous les 300 usages ou 3 mois
Procédure d'entretien
• Lubrifiez le clapet antiretour interne en retirant le piston de pompage et
le clapet antiretour interne. Déposez 2 ml d'huile dans la partie filetée
du clapet antiretour et réassemblez. Lorsque le piston de pompage sera
actionné, l'huile sera pompée à travers le clapet antiretour interne.
• Lubrifiez le clapet antiretour externe en retirant le clapet antiretour
externe. Déposez 2 ml d'huile dans la partie filetée du clapet antiretour et
réassemblez. Lorsque le piston de pompage sera actionné, l'huile sera
pompée à travers le clapet antiretour externe.
• Lubrifiez les raccords Ralston Quick-test™ en déposant 2 ml d'huile à
l'intérieur des raccords.
• Lubrifiez les joints toriques du piston de pompage et du piston de réglage
micrométrique avec du lubrifiant à la silicone.
Manutenção .....................................19
Manutenzione ..................................19
Wartung ...........................................20
17