Entering
Calibration/Adjustment
and Linearization Weights
Purpose
To enter values for the calibration/
adjustment and linearization weights.
Notes on Settings
– The service mode must be active for
entering the linearization weights under
menu items 1.18.2 through 1.18.5.
Diagram of Menu for Entering Calibration/Adjustment and Linearization Weights
1. 18.
1. 18. 1
1. 18. 2
1. 18. 3
1. 18. 4
1. 18. 5
16
– Enter calibration/adjustment and lin-
earization weights in the Setup menu
under "COMSPEC". These settings are
configured in the corresponding
numeric menu under menu item 1.18.
– The service mode is not required for
configuring menu item 1.18.1, "Exter-
nal user-defined weight."
Entering Calibration/Adjustment and Linearization Weights
Enter external user-defined calibration weight (service mode not required)
Enter linearization weight 1
Enter linearization weight 2
Enter linearization weight 3
Enter linearization weight 4
Procedure
– Activate the service mode (required only
for entering linearization weights)
– Select the weighing platform
– Enter the external user-defined weight
under menu item 1.18.1
– Enter external linearization weights
under menu items 1.18.2 through
1.18.5