3.
Press the ENTER key to display the measurement stored inside the selected
memory location, the instrument displays a screen similar to the one below
4.
Press the ESC key to go back to previous screen and press the ESC key again to
exit the memory management
6. INSTRUMENT RESET AND DEFAULT PARAMETERS
�CAUTION
Before carrying out the instrument's reset save all data relative to the measurements effected
by downloading them to a pc.
1.
When the instrument is off press the RCL/CLR key
2.
Keeping down the RCL/CLR key, press the switch on key. The instrument emits a
short sound tone showing all display segments for approx. 1 second. Then it
emits a second short sound tone displaying the screen beside for approx. 3
seconds
�CAUTION
The HARD RESET procedure deletes all data previously stored and the parameter DST resumes
its default value (1 m or 3 ft)
7. INSTRUMENT CoNNECTIoN To PC
The instrument can be connected to a PC by means of the serial port or USB and opt insulated
cable provided along with the software package. First it's necessary to select the COM port used
for the transmission and the correct baud rate (9600 bps). To set these parameters install the
software and consult the help on line.
The selected port shall be free of any other device or application such as mouse, modem, etc.
�CAUTION
Optical port emits Laser radiations , Don't locate beam at eye level. Class 1M laser apparatus
according to EN 60825-1
(Measurement stored in the selected memory
location)
(Values of interfering voltage at measurement)
19