Operation cont.
Fuse Replacement
WARNINGS:
To avoid electric shock, remove the test leads from
the meter before removing the battery/fuse cover.
1. Lift up the tilt stand.
2. Loosen the Phillips
screws on the battery/fuse
cover.
3. Remove the battery/fuse
cover.
4. Gently remove fuse and
install new fuse into the
holder.
5. Always use a UL recognized
fuse of the proper size and
value: 10A/600V (5 x 20mm)
fast blow.
6. Install the battery cover and
tighten the screws
WARNING:
To avoid electric shock, do not operate meter until
the battery/fuse cover is securely fastened to the meter.
WARNING:
When replacing the battery or fuses, be sure to secure
the battery compartment door firmly to maintain the waterproof and dust
proof integrity of the meter. Loose or overtightened screws, or an improperly
seated o-ring may compromise the meter's water and dust ingress protection.
Fuse
15
Specifications
Accuracy is given at 65°F to 83°F (18°C to 28°C), less than 70% relative humidity
Function
Range
4.000V
AC Voltage
40.00V
400.0V
600V
Input Protection:
600V AC RMS or 600V DC
AC Response:
50 to 60Hz
Function
Range
400.0mV
DC Voltage
4.000V
40.00V
400.0V
600V
Input Protection:
600V RMS or 600V DC
Function
Range
4.000A
AC Current
10.00A
Overload Protection:
10A/600V Fuse
AC Response:
50 to 60Hz
Function
Range
4.000A
DC Current
10.00A
Overload Protection:
10A/600V Fuse
Resolution
Accuracy ±
(% of reading + digits)
1mV
10mV
±(1.5% +10 digits)
0.1V
1V
Resolution
Accuracy ±
(% of reading + digits)
0.1mV
±(0.8% +8 digits)
1mV
10mV
±(1.2% +2 digits)
0.1V
1V
±(1.2% +5 digits)
Resolution
Accuracy ±
(% of reading + digits)
1mA
±(3.0% +5 digits)
10mA
Resolution
Accuracy ±
(% of reading + digits)
1mA
±(3.0% +5 digits)
10mA
16