Place the clamp around the active conductors (Phase conductors
and Neutral). and release the trigger; check that the clamp is properly
closed. Read the measurement on the display unit.
The reading can be filtered to reflect only fundamental by a long press
on the HOLD key, giving an indication of the impact of the harmonics.
4.3. AC AND DC VOLTAGE MEASUREMENT
Set the switch to V~ for an AC voltage measurement and to V= for a
DC voltage measurement.
Connect the red test cord to the "+" input terminal and the black test
cord to the «COM» input terminal.
Then place the probe tips in contact with the points where the AC
voltage is to be measured.
Read the result on the display unit.
28