Longevity Use; Notes; Accessories; Troubleshooting - VISIOMED ThermoFlash LX-361T Manual De Utilización

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 20

LONGEVITY USE

The THERMOFLASH® LX-361T is guaranteed for 40.000 readings.
EMC DATA
Guidance and manufacturer's declaration - electromagnetic immunity - for equipment and sys-
tem that are not life-supporting. The model LX-361T is intended for use in the electromagnetic envi-
ronment specified below. The customer or the user of the model LX-361T should assure that it is used
in such an environment.
EN
Immunity
60601
Compliance
test
test
level
level
Conducted
Portable and mobile RF communications equipment should be
3 Vrms
N/A
RF
used no closer to any part of the model LX-361T, including cables,
than the recommended separation distance calculated from the
150 kHz
equation applicable to the frequency of the transmitter.
IEC
to 80
Recommended separation distance:
61000-4-6
MHz
d =
3 V/m
Radiated RF
3 V/m
d =
80 MHz
IEC
to 2.5
61000-4-3
GHz
where P is the maximum output power rating of the trans-
mitter in watts (W) according to the transmitter manufacturer
and d is the recommended separation distance in meters (m).
Fields strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, a should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of equipment marqued
with the following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic is affected by absorption and reflection from
structures, objects and people.
a. Fields strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the model LX-361T is used ex-
ceeds the applicable RF compliance level above. The model LX-361T should be observed to verify normal opera-
tion. If abnormal performance is observed, additional measures may be necessary such as reorienting or relocating
the model LX-361T.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
Recommended separation distances between portable and mobile RF communications equip-
ment and the equipment or system - for equipement and systems that are not life-supporting.
The model LX-361T is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the model LX-361T can help prevent elec-
tromagnetic interference by maintaining a minimum distance between portable and mobile RF com-
munications equipment (transmitters) and the model LX-361T as recommended below, acording to the
maximum output power of the communications equipment.
24
Visiomed - LX-361T - USER MANUAL - 012017
Electromagnetic
environment - guidance
[ 3.5
] √
P
80 MHz to 800 MHz
E1
[ 7
] √
P
800 MHz to 2.5 GHz
E1
b
Rated maximum output
of transmitter
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above the recommended separation distance d in me-
ters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagations is affected by absorption and
reflection from structures, objects and people.
IX. NOTES
- The glass protecting the lens is the most important and most fragile part of the thermometer, so be
careful with it.
- Clean the glass with a cotton cloth, dampened with water or 70% alcohol.
- Only use the recommended batteries, do not recharge non-rechargeable batteries, do not burn them.
a
- Remove the batteries if the thermometer is not to be used for a long period.
- Do not expose the thermometer to direct sunlight or to water.
- A shock may damage the thermometer.
X. ACCESSORIES
User manual, Warranty card, Base, Batteries (LR03).
XI. TROUBLESHOOTING
I
f you have one of the following problems while using your THERMOFLASH® LX-361T, please refer to
this breakdown service guide to help resolve the problem. If the problem persists, please contact our
customer service at +33 892 350 334.
- The screen is showing temperatures higher than 95°C: the temperature is in Fahrenheit. Change the
measurement unit to Celsius by pressing the Program setting button to get F1 function key.
- The screen is showing temperatures lower than 32°C (89.6°F): to take the internal temperature of
your body, choose the BODY function. If you are in SURFACE TEMP or ROOM mode, the 32°C temperature
displayed is showing the external temperature that your body releases.
- The screen is showing the message HI: the temperature taken is higher than the range of the mode
selected, either higher than 42.9°C (109.2°F) in BODY mode or 60°C (140°F) in SURFACE TEMP mode.
- The screen is showing the message LO: the temperature taken is lower than the range of the mode
selected, either lower than 32°C (89.6°F) in BODY mode or 0°C (32°F) in SURFACE TEMP mode.
Separation distance according to
frequency of transmitter m
80 MHz to 800 MHz
[ 3.5
d =
] √
P
E1
0.12
0.38
1.2
3.8
12
Visiomed - LX-361T - USER MANUAL - 012017
800 MHz to 2.5 GHz
[ 7
d =
] √
P
E1
0.23
0.73
2.3
7.3
23
25

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido