CALIBRATION

One point calibration.

1. Under normal weighing mode, press
"SLEEP".
2. Press the
tray stable.
3. Display: xxx, clue to load xxx g poise, now you can press
load point, you can load on the poise which you have choose, press the
key to confirm and keep it stable.
4. If the weight is not correct after calibration, you can try two point calibration.
Note: The two calibration methods are based on unit gram.

Two point calibration.

1. Enter to calibration mode, press the
"SLEEP"
2. Press the
3. The display will show "F-XX" to choose capacity, press the
and press the
4. Display will show: d=x to choose division, press the
key to confirm.
5. Display will show: CAL- to catch zero, empty the tray and keep it stable, then
press
key to confirm, display will view "--------" catching the zero.
6. Display will show: xxx, clue to load xxx g poise, this is one point calibration,
after load the poise, press
key to catch zero, the display will show "--------" and keep the
key again to enter into calibration mode.
key to confirm.
key to confirm, display will view "-------".
key for 2 seconds until display shows
key for 2 seconds until display shows
to enter division and
to choose the
key to choose
7