Icon displayed
If any of the following marks is displayed, take the specified
action, then repeat the operation.
Icon displayed

KEY TO DISPLAY ICONS

Condition/Cause
• Appears when the
blood-pressure could
not be measured
accurately.
The pressure is
281mmHg or higher.
Appears when the
batteries are running low.
Detected problem in the
device.
- Eng 20 -
Condition/Cause
See page 15.
See page 15.
Lit
Measurement is under way.
Blinking
Pulse detected.
"Er" is displayed when the pulse value
is below 40 or above 181.
Corrective action
Apply cuff properly and
measure again.
If the pressure does not
automatically decline
during measurement,
immediately press the
"START/STOP" switch
and turn the unit off. Take
the measurement again.
Replace the batteries.
Contact the vendor.