Be sure to follow the installation, operating, maintenance and servicing instructions given in the manuals supplied.
1)
USA: The installation must be in accordance with the National Electrical Code
ANSI / ISA-RP 12.6.
Canada: The installation must be in accordance with the Canadian Electrical Code
2)
No connection to any device that uses or generates in excess of 250Vrms or 250Vdc.
3)
USA: The apparatus must be connected to a suitable ground electrode per National Electrical Code
Article 504 or 505. The resistance of the ground pad must be less than 1 ohm.
Canada: The apparatus must be connected to a suitable ground electrode per Canadian Electrical Code
18. The resistance of the ground pad must be less than 1 ohm.
4)
Only the Sartorius cable type YCC01-0052Mxx (XX = 10 for 10m, 20 for 20m; 30 for 30m) may be used. Maximum
cable length: 100ft (30.5m). The equipment operator is responsible for any non-Sartorius cables used.
5)
If the Ex-Link Converter YCO14-Y is installed in Zone 2, the installation must be done in such a way that the
pollution degree 2 of the YCO04-Y board is guaranteed.
6)
For use in Division 2: The Ex-Link Converter YCO14-Y must be connected via an USB cable to a nonincendive
apparatus with nonincendive field wiring output. The input parameters of the YCO14-Y are
Vmax = 17.0V, Imax = 2.0A, Pmax = any, Ci = 5.7μF, Li = 0mH or Vmax = 16.0V, Imax = 2.5A, Pmax = any,
Ci = 5.7μF, Li = 0mH (Entity parameters).
For use in Zone 2:
The Ex-Link Converter YCO14-Y must be connected via an USB cable to a apparatus with intrinsically safe (Ex ic)
output. The input parameters of the YCO14-Y in type of protection Ex ic IIB are
Ui = 17.0V, Ii = 2.0A, Pi = any, Ci = 5.7μF, Li = 0mH or Ui = 16.0V, Ii = 2.5A, Pi = any, Ci = 5.7μF, Li = 0mH (Entity
parameters).
7)
In Zone 2 the „USB Power" port may not be used.
8)
If the cable contains no Ex ic circuits, the cable must be protected against unintended release by use of Sartorius
„Strain Relief for USB Cord" type YSR01.
9)
The equipment shall be installed in such a way that it is protected against the entry of solid foreign objects or
water capable of impairing the safety of the apparatus. Reduce the risk of mechanical damage to a minimum.
10) Exposure to UV radiation is not allowed!
11) Use equipment only in the temperature ranges indicated. Avoid exposing the equipment to heat.
12) The connecting cable of the display unit must be prevented against damage and stress caused by strain.
13) If the equipment does not operate properly, unplug it immediately from line power (mains supply)!
14) Avoid generating static electricity. Use only a damp cloth to wipe down the equipment. The equipment operator
shall be responsible for preventing any risks caused by static electricity.
15) Keep chemicals and other agents, which can corrode the housing seals and cable sheaths, away from the
equipment. These agents include oil, grease, benzene, acetone and ozone. If you are not sure about the safety of a
certain substance, please contact the manufacturer.
16) At reasonable intervals, have your equipment installation checked for proper functioning and safety by a trained
and certified technician.
17) Prior to opening the equipment, disconnect the power supply or make sure that there is no potentially explosive
atmosphere or any other explosion hazard in the surrounding area!
18) WARNING: SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY.
If your equipment needs to be repaired, use only original spare parts supplied by the manufacturer!
19) Any tampering with the equipment by anyone, other than repair work done by authorized Sartorius service
technicians, will result in the loss of EX conformity and in the forfeiture of all claims under the manufacturer's
warranty. Only authorized specialists may open the equipment.
20) Modifications, including those to be carried out by Sartorius employees, may be permitted only after the express
written authorization has been obtained from Sartorius.
2015-10-10
Dr. D. Klausgrete
Conformidad y licencias
®
, NFPA 70, Article 504 or 505 and
®
, Section 18.
Control Drawing
EVO1X.....-..... / LAB1X.....-..... + YCO14-.
2021459
01
Revision
Instrucciones de manejo EVO1X | LAB1X
®
, NFPA 70,
®
, Section
2
2
Sheet
of