• Installation
Before connecting the aspirator to the piping of the cen-
tralized system, ascertain that aspiration piping is clean as
heavy debris could damage the appliances.
Connect the PVC light grey aspiration tube (2b) (supplied
with the machine) to the 50 mm Ø tube-holder (2) ("aspirated
fl uid inlet").
The other end of the same tube must be connected to the
aspiration piping (3) coming from the surgeries (page 24).
The black heat resistant exhaust air pipe (4b), fi tted with a
metal spiral, must be connected to the 50 mm Ø tube-holder
(4) ("exhaust air outlet").
Connect the other end of the pipe to the antibacterial fi lter
(5), passing preferably through a silencer (5a) supplied with
the aspirator. The hot air coming from the antibacterial fi lter
must be conveyed outside.
Connect the 18 mm Ø tube-holder (6) to the liquid drain pipe.
In the version fi tted with Hydrocyclone the aspirator is drai-
ning liquids by gravity, therefore these can never be drained
towards the top. The 10 mm Ø tube-holder (8) ("emergency
drain") must be connected to the liquid drain pipe.
The piping connecting the machine to the aspiration and
draining system should be fl exible to damper the small
vibrations produced by the aspirator.
The aspiration piping should be run in the fl oor and at a point
near the aspirator it should rise about 30 cm to reach the
tube-holder (2) (fi g. A and B, page 91).
If Turbo-Smart is installed on a lower
fl oor than the dental units, the aspiration
piping must not enter the centrifugal
separator perpendicularly.
Rather, place a few metres of piping
horizontally on the same level of the
machine and then install it at the centri-
fugal separator inlet which is on a higher
level (use a fl exible pipe) (fi g.B, page
91). The aspirated fl uid is reaspirated by
Turbo-Smart.
After the installation, connect the power
cable to the mains according to the
regulation EN 61008-1.
Finally connect the low voltage users line:
the clamps 19 and 20 of the circuit AC
80 (fi g. C "AC 80 wiring diagram", page
92) to the users line (12) (det.12, page 24
and fi g. F, page 95) which is feeding all
the dental units of the same system with
connections in parallel.
Ascertain that the contact on the dental
unit is clean (free contact without vol-
tage).