Use Instructions - 3A HEALTH CARE MINIASPEED Battery Evo Plus Manual De Instrucciones

Ocultar thumbs Ver también para MINIASPEED Battery Evo Plus:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

The MINIASPEED BATTERY EVO PLUS aspirator is a professional portable device suitable for use in ambulances,
specific for secretion removal. It has a vacuum regulator, a vacuum gauge and 1000 ml container with a protection
device to prevent liquid from entering the suction pump, which cuts off the aspiration flow.
It has lifetime lubrication and is easy to handle, simple to use, reliable, strong and silent. This model has a 12 V
rechargeable battery; when connected to the ambulance mounting bracket it not only operates on the ambulance's
battery but also recharges. The MINIASPEED BATTERY EVO PLUS aspirator has the following 3A accessories:
1000 ml container with protection device, power supply cord with multi-voltage power supply unit, 24 cm long
sterilisable silicone connection tube, 130 cm long sterilisable silicone connection tube, sterile, disposable cannula,
sterile, disposable manual flow regulator, disposable bag and disposable antibacterial filter.
N.B.: Only use original 3A accessories.
This is a medical device and must be used by qualified staff. It must be operated as indicated in this user
instruction manual. It is important for the operator to read and understand the information on use and
maintenance of the unit. If you have any questions, contact your stockist. MICROBIAL CONTAMINATION:
in the case of illnesses with a risk of infection or microbial contamination, the accessories should be
thoroughly cleaned and sterilised after each use.
The manufacturer has made every effort to ensure that all its products are of the highest quality and are reliable.
Nevertheless, as for all electrical appliances, fundamental safety standards must be observed:
• Children and persons who are not self-sufficient may only use the unit under the strict supervision of a
responsible adult who has read this manual.
• The device must always be used by specifically trained staff who have read this manual.
• Never use adapters for supply voltages different to the voltage shown on the data plate on the back of the
unit. Keep the cord away from hot surfaces.
• The unit should not be used in the presence of inflammable anaesthetic mixtures with air, oxygen or nitrous
oxide.
Never handle the supply cord plug with wet hands or use the unit when taking a bath or a shower.
Never leave the unit close to water, do not submerge it in any liquids. If by chance it falls into water,
pull the plug out of the socket immediately before recovering it. Do not use the unit if the plug or
supply cord are worn or wet (send it immediately to your stockist).
Although the device is protected against sprays of water, do not use it in the rain.
• Only authorised personnel may perform maintenance and/or repair work. Unauthorised repairs annul the
guarantee.
• Ensure that the connections and container closing are performed carefully to prevent suction losses.
• Do not tip the container over while connected to the unit when operating, as liquid may be sucked inside the
appliance causing pump damage. If this happens, switch the aspirator off immediately and empty and clean
the container. Send the appliance to your stockist.
• When the protection device intervenes aspiration is interrupted; empty the container and perform the cleaning
operations.
• The cannula and the manual control of the aspirated flow are sterile, disposable products: They must be
replaced after every application.
• Check the use-by-date on the original packaging of the cannula and check the integrity of the sterile packaging.
If expired and/or deteriorated, replace it.
• The disposable antibacterial filter must be replaced after every application.
• Never use the battery charger with other appliances or for uses other than that established by this manual.
Never use MINIASPEED BATTERY EVO PLUS with other power supply units.
• In view of their length, the power cord and connection hose could constitute a strangulation hazard.
• Use of the device in ambient conditions other than those specified in the manual may seriously impair its
safety and technical characteristics.
• In the event of aspiration without the container and/or antibacterial filter, or if it is suspected that substances
have entered the aspiration circuit, contact your stockist at once.
• Never modify the device's mounting bracket or the bracket installed in the ambulance as this may impair the
safety of both the device and its installation environment.
• Always use the device in the vertical position on an unobstructed, stable, flat surface.
• The lifetime of the device itself is 5 years, and that of the container and silicone tubes 1 year or 30 sterilisation
cycles. However, the lifetime may vary depending on the operating environment.
Each time before use, ensure that all the accessories are perfectly clean according to the instructions indicated
in the "CLEANING AND DISINFECTION OPERATIONS".
GB
IMPORTANTS WARNINGS

USE INSTRUCTIONS

18

Publicidad

Tabla de contenido
loading

Tabla de contenido