INSTALLATION
2.2. INSTALLATION TO THE WALL
Place the Wall Support and the Timer at the desired location on the wall and mark the holes in the wall,
1
using the Wall Plate as a template
Drill the four pilot holes:
2
- for the Wall Support (with a 9 mm drill bit)
- for the Timer (with a 8 mm drill bit).
Position the Timer over the four pilot holes, hav-
3
ing routed the line cable through the Timer knock
out and having routed additional cable from the
Timer to the Wall Support.
B-6
additional cable
Timer-Wall Support
line cable
(mains)
Printed on: 22 Mär 2012, 09:28:10 am; Printed by: TOBIAS BAUER
... and the template on page BZ-2 for the Timer.
ORALIX AC SYSTEM
WARNING
Do not use the Timer as
a template ! Electronic
components could be
irreparably damaged !