INSTALLATION

2.6. MOUNTING OF THE TUBEHEAD

Before mounting the Tubehead remove very care-
1
fully the security binding of the Folding Arm and
guide the arm slowly to the horizontal position
opposing by hand to the springs action directed
upwards.If the Folding Arm is released without
Tubehead and without opposing by hand to the
springs action, it is possible to damage the
springs themselves.
2
Partially loosen the screw (a) ...
and remove cover plate (b).
3
Shift upwards handle (c) and remove "U" shaped
retaining clip (d).
4
Remove clear plastic tube from Tubehead con-
tact assembly.
Lightly grease the Tubehead contact pin (g) and
inner part of Folding Arm.
WARNING
Do not grease electrical connections on top of
Tubehead contact pin and into Folding Arm.
B-14